10.3 Adjusting Measurement Functions
Table 10.2 Adjustment Setpoints of Measurement Functions
DC 100 mV
DC 1 V
DC 10 V
DC 100 V
DC 20 mA
DC 100 mA
400
Ω
AC 1 V
AC 10 V
AC 100 V
AC 300 V
100 mV
1 V
10 V
100 V
20 mA
100 mA
380
Ω
1 V/50-60 Hz
10 V/50-60 Hz
100 V/50-60 Hz
300 V/50-60 Hz
–
–
–
–
–
–
0
Ω
0 V
0 V
0 V
0 V
Range
Adjustment Setpoint
*2
CAL FS
CAL 0
Remarks
Three-wire connection
*2: Apply the reference input signals listed above from the calibration standard.
TIP
• You can also select only the range in need of readjustment to adjust it separately.
• Always make zero-point (0) adjustments together with full-scale (FS) adjustments.
10.3.1 Adjusting DC Voltage and DC Current Ranges
Step 1: Press the
key while simultaneously holding down the
key. The LCD shows
.
Step 2: Pressing the highest-order
key causes the LCD to show
.
Step 3: Pressing the
key enters the measurement CAL mode.
The
indicator blinks on the LCD and the
symbol
appears. The calibrator is now ready for the defining the CAL
FS setpoint of measurement functions.
CAL-mode Operation Keys and Display Indications
10.3 Adjusting Measurement Functions