Divided Output Function (
n/m)
The divided output function (
n/m) outputs a value n/m times the
setpoint of a voltage, current, resistance, thermocouple or RTD signal.
Thus, the output value is defined as:
Output value = Main setpoint
×
(
n/m)
Keys and labels related to divided output function (
n/m
)
For details on how to set the sourced signal level of each range, see
Sections 4.2, “Sourcing DC Voltage, DC Current or SINK Current Sig-
nal, to 4.4, “Sourcing Thermocouple (TC) Signal.” Follow the steps
shown below with the calibrator output turned off.
Step 1: When the setting of the sourced signal level of each range is
complete, follow step 2 and later steps.
Step 2: Using each pair of
or
keys, set the main setpoint.
Step 3: Press the
key to enter the divided output (
n/m) mode.
The LCD shows
. The higher-order two digits repre-
sents the value of
n and the lower-order two digits the value of
m.
Step 4: Using a pair of
or
keys, set the value of
m. The variable
m can be set to a value from 1 to 19.
Step 5: Using a pair of
or
keys, change the value of
n. An
output value
n/m times the main setpoint can be obtained ac-
cording to the setpoint of
n. The variable n can be set to a
value from 0 to
m.
4.6 Divided Output Function (
n/m)