7.1
Selecting the Measurement Mode
Procedure
1.
Press
SHIFT+V/I (MODE)
in the MEASURE section of the front panel to display
the Measure Mode menu.
2.
Press the soft key corresponding to the desired measurement mode.
Explanation
Select from the modes below.
•
Off
Measurement OFF:
Does not perform measurements.
•
Fixed
Fixed function mode:
Select a measurement function (see section 7.2, “Selecting
the Measurement Function”) and perform the measurement.
•
Auto
Auto function mode:
When the source function is switched, the measurement
function automatically switches so that it is not the same
function as the source function. However, if the limiter is
activated, the GS820 measures the same function as the
source function.
•
V-Meter
Voltmeter mode:
Generates 0 nA and measures the voltage.
•
I-Meter
Ammeter mode:
Generates 0 mV and measures the current.
•
R-Meter
Resistance meter mode: Generates a fixed current at each resistance range,
measures the voltage, and calculates the resistance as a
result.
<<Corresponding Command Mnemonic>>
[:CHANnel<n>]:SENSe[:STATe] 1|0|ON|OFF
[:CHANnel<n>]:SENSe:MODE FIXed|AUTO|VMETer|IMETer|RMETer
Chapter 7
Measurement
7-1
IM 765601-01E