5.8
Selecting the Response Mode
Procedure
1.
Press
CONFIG
in the SOURCE section of the front panel to display the CONFIG
menu.
2.
Press the
Response
soft key to display the Response menu.
3.
Press the soft key corresponding to the desired response mode.
Explanation
Response Mode
Select an appropriate response mode according to the DUT used or the application
objective.
Normal: The response time is short, but the GS820 may become unstable and oscillate
when a highly inductive or capacitive load is connected.
Stable: This mode is robust to inductive and capacitive loads, but the response time is
longer.
<<Corresponding Command Mnemonic>>
[:CHANnel<n>]:SOURce:RESPonse NORMal|STABle
5-14