<2. Installation, Wiring and Piping>
2-6
IM 11T03E01-51E
2.2.2
Mounting the Converter
l
Panel Mounting
Unit: mm
F0209.ai
Fixing screws (2 pcs)
Bracket
Converter
Insert the converter in
the panel cutout before
attaching the brackets.
Panel
Dimension of Panel Cutout
185
216
160
(173.5)
(83)
(3)
139
+1
0
Figure 2.8
Panel Mounting
2.3 Piping
CAUTION
In the case of replacement range, H
2
in CO
2
or Air in CO
2
, the output happen to be fluctuated by
pressure loading to detector. In order to reduce the effect, flowmeter is to be set upper side of
detector so that the pressure loading to the detector is going to be about atmospheric pressure.
The piping connected to the analyzer comprises a line that feeds the gas under measurement
to the detector, a line that returns (or releases to the atmosphere) the gas exhausted after
measurement, and lines that carry the zero and span calibration gases. The type of piping
chosen depends on the composition of the gas under measurement, its pressure, the amount of
dust mixed, and the response (dead time). It is advisable however to use stainless steel piping of
sizes from 6 mm and 4 mm in outer and inner diameters up to JIS 15 A (21.7mm in outer dia.).
Refer to the following instructions when connecting the analyzer piping.
l
Figure 2.10, "System Configuration," shows an example of piping. Connect the lines
securely to ensure that there is no gas leakage in the system.
l
Make sure the pressure of the sample gas is no greater than 0.5 MPa when measured at
the detector inlet. If the pressure is too high, use a pressure reducing valve to regulate the
pressure to a normal level. If the pressure is too low, it must be raised using a pump.
l
If the gas under measurement contains dust, mist or moisture, such impurities must be
removed. Install a filter, mist separator or dehumidifier to remove these impurities from the
gas.
l
In order to return the gas exhausted after measurement, the difference in pressure between
the detector inlet and the point of returning the gas must be 0.5 kPa minimum. Choose
piping with a large inner diameter to minimize pressure loss in the return gas line.
l
ALWAYS install a stop valve at the point where the gas is sampled (or returned).
l
The pressure transmitter for compensating pressure is designed to detect the pressure
inside the detector; install a pressure lead pipe as close to the detector as possible
(preferably, no more than 0.5 m away).
Summary of Contents for GD402G /M1
Page 21: ...Blank Page ...
Page 37: ...Blank Page ...
Page 81: ...Blank Page ...
Page 115: ...Blank Page ...
Page 159: ...Blank Page ...
Page 167: ......
Page 169: ......
Page 171: ...Blank Page ...