11.2 Inspection In Case of Failure
11.2.1
Inspecting the Analyzer in an Alarm Status
The GD402 analyzer outputs an ALARM signal through terminals 16 and 17 if a failure is
detected by the self-diagnostic function during measurement. In addition, the <ALARM> lamp
on the operation panel comes on and the message field shows an alarm number. The analyzer
outputs no contact signal, however, if the alarm number is [ALM.09]. If a failure (alarm) is
detected, take the corrective action given in Table 11.1.
Table 11.1
Corrective Actions when an Alarm Status Is Detected
Message
Shown
Failure Type
Diagnostic Conditions
Corrective
Actions
Message via
Communication
ALM.01
High/low limit alarm
against physical
density
Physical density ≤ Low-limit-of-
physical-density alarm set-point
Physical density ≥ High-limit-of-
physical-density
Change the high/low
limits.
OUT OF DENSITY
ALM.02
High/low limit alarm
against compensated
density
Compensated density ≤ Low-limit-of-
compensated-density alarm set-point
Compensated density ≥ High-limit-of-
compensated-density alarm set-point
Change the high/low
limits.
OUT OF COMP
DENSITY
ALM.03
High/low limit alarm
against specific gravity
Specific gravity ≤ Low-limit-of-specific
gravity alarm set-point
Specific gravity ≥ High-limit-of-specific
gravity alarm set-point
Change the high/low
limits.
OUT OF SP GR
ALM.04
High/low limit alarm
against calorific value
Calorific value ≤ Low-limit-of-calorific-
value alarm set-point
Calorific value ≥ High-limit-of-calorific-
value alarm set-point
Change the high/low
limits.
OUT OF CALORY
ALM.05
High/low limit alarm
against molecular
weight
Molecular weight ≤ Low-limit-of-
molecular weight alarm set-point
Molecular weight ≥ High-limit-of-
molecular weight alarm set-point
Change the high/low
limits.
OUT OF
MOLECULE
ALM.06
High/low limit alarm
against concentration
Concentration ≤ Low-limit-of-
concentration alarm set-point
Concentration ≥ High-limit-of-
concentration alarm set-point
Change the high/low
limits.
OUT OF CONC
ALM.07
Error in pressure input
range
Zero point -3% ≥ pressure
Span point +5% ≤ pressure
Pressure ≤ 0.1 kPa
Check the sample-gas
pressure and the
pressure range.
OUT OF GAS
PRESS
ALM.08
Abnormal sample gas
temperature
-25 °C ≤ sample gas temperature
≤ 80°C
Use the sample gas
within the tolerance
limits.
OUT OF GAS
TEMP
ALM.09
Battery failure
Contact Yokogawa
service personnel.
LOW BATTERY
ALM.10
Error in calibration
(zero and span points)
-0.3000 ≤ zero calibration value
≤ 0.3000
0.50000 ≤ span calibration value
≤ 1.5000
Redo calibration.
ILLEGAL CALIB