<8. DENSITY / CALIBRATION PROCEDURE>
8-3
IM 11T03E01-51E
8.2.2
Setting Calibration Data
To set calibration data, press the [*] key in the measurement mode, (when password is selected,
enter the password (XXX), ) and select the parameter [*CAL.DT]. See Figure 5.4 for the process
at the setting level.
(a) Zero-point density [*Z_DNS]
Configurable range : 0.0000 to 6.0000 (kg/m
3
)
0.00000
to
0.40000
(lb/ft
3
)
(b) Span-point density [*S_DNS]
Configurable range : 0.0000 to 6.0000 (kg/m
3
)
0.00000
to
0.40000
(lb/ft
3
)
(c) Output hold value during calibration [*C_HLD]
•
Disable:
0
•
Enable:
1
•
Enable (preset value):
2
(d) Setting of preset value [*PR.SET]
Configurable range:
-10.0
to 110.0
(Applies only if item (c) above is set to “2”.)
[NO]
[ENT]
*RANGE
[NO]
X.XXXX
*Z_DNS
*ALARM
*CAL.DT
X.XXXX
*S_DNS
X
*C_HLD
XXX.X
*PR.SET
[ENT]
[ENT]
[ENT]
F0803.ai
[YES]
*
2
*
2
*
2
*
2
*
3
*1:
See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*3: This step is skipped if the parameter [*C_HLD] is set to a value other than "2".
*
1
(Setting of Zero-point density)
(Setting of preset value)
(Setting of Output hold value during calibration)
(Setting of Span-point density)
Figure 8.2
Calibration Parameter Setting Mode