2. Installation, Piping, and Wiring
67
TI 11B08A01-01E
2.3.2
Recommended Cables
CAUTION
(C) to (L) cables:
Use heat-resistant cables with maximum allowable temperature of 80°C or higher.
(M) and (N) cables:
Required maximum allowable temperature of the cables depends on the temperature class
of the instrument and the actual ambient temperature. Use cables with maximum allowable
temperature shown in the table below.
Temperature Class
T1, T2
T3
T4
Maximum allowable
temperature of cable
Ambient temp.
Up to 30°C Up to 35°C Up to 40°C
60°C
31 to 36°C 36 to 40°C 41 to 45°C
65°C
37 to 43°C 41 to 45°C 46 to 50°C
70°C
44 to 50°C 46 to 50°C
75°C