![YOKOGAWA FLXA402T User Manual Download Page 31](http://html.mh-extra.com/html/yokogawa/flxa402t/flxa402t_user-manual_911266031.webp)
<3. Converter menu>
3-8
IM 12A01G01-03EN
1st Edition : Mar. 25, 2021-00
3.4.2
Lang.(Language)
3.4.3
Initialization
If you need to reset to factory default, follow this procedure. In a usual operation, you don’t need
the initialization.
CAUTION
Once you initialize the converter, every data of, language, password, device setting is cleared
and reset to default, excluding date/time.
If you want to initialize after configuring some settings, store the data into the memory of
FLXA402T, if necessary. See
To connect a sensor to FLXA402T and display the measurement or mA output, you need to set
up a communication network with sensors or assign each sensor-module address to HMI.
l
How to reset parameters
1.
On Home/Main screen, tap to go to Converter menu.
2. Press
“Setting”.
3. Press
“Initialization”.
4.
On “Load factory defaults”, find the sensor module type in use and select.
Connection number is either “1-1” or “1-1+2-1”. Select an appropriate connection to suit
your sensors.
Table 3.1
List example of Load factory defaults
TB
TB + PH
CL
TB + SC
TB + TB
CL + PH
CL + TB
CL + SC
CL + CL
5.
Confirm the load selection you choose on the step 4. (above) is displayed and tap
“Execute”.
(1)
(2)
(3)
(4)
(5)
Figure 3.7
How to initialize parameters