![YOKOGAWA FLXA21 User Manual Download Page 40](http://html.mh-extra.com/html/yokogawa/flxa21/flxa21_user-manual_911262040.webp)
<6. Function block parameters and Methods>
6-14
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
Table 6.2 Contents of Classic DIAGNOSIS
Octet Bit DIAGNOSIS Mnemonic
Description
Remarks
1
0 DIA_HW_ELECTR
Hardware failure of the electronic
Refer to Table of Device
status.
1 DIA_HW_MECH
Hardware failure mechanics
Refer to Table of Device
status.
2 DIA_TEMP_MOTOR *2
Motor- temperature too high
3 DIA_TEMP_ELECTR *2 Electronic temperature too high
4 DIA_MEM_CHKSUM
Memory error
Refer to Table of Device
status.
5 DIA_MEASUREMENT *2 Failure in measurement
—
6 DIA_NOT_INIT *2
Device not initialized (No self calibration)
—
7 DIA_INIT_ERR *2
Self calibration failed
—
2
0 DIA_ZERO_ERR
Zero point error (limit position)
Refer to Table of Device
status.
1 DIA_SUPPLY *2
Power supply failed (electrical,
pneumatic)
—
2 DIA_CONF_INVAL *2
Configuration not valid
—
3 DIA_WARMSTART *1
New start-up (warm startup) carried out. Should be set after power-on
or after FACTORY_RESET =
2506 has been executed.
4 DIA_COLDSTART *1
Restart (cold startup) carried out.
Should be set after
FACTORY_RESET = 1 has
been executed.
5 DIA_MAINTAINANCE *2 Maintenance required
—
6 DIA_CHARACT *2
Characterization invalid
—
7 IDENT_NUMBER_
VIOLATION *2
Set to 1 (one), if the IDENT_Number
of the running cyclic data transfer and
the value of Physical Block IDENT_
NUMBER_SELECTOR parameter do
not correspond. If IDENT_NUMBER_
SELECTOR = 127 (adaption mode) then
the DIAGNOSIS bit IDENT_NUMBER_
VIOLATION is cleared / not set.
—
3
0-7 Reserved *2
Reserved for use within the PNO
—
4
0-6 Reserved *2
Reserved for use within the PNO
—
7 EXTENSION_AVAILABLE More diagnosis information is available
—
*1:
This bit is Off 10 seconds after On.
*2:
Not available for FLXA21.
n
Device status
Device setting status and failures of FLXA21 are indicated by using parameter DEVICE_
STATUS_1, DEVICE_STATUS_2, DEVICE_STATUS_3, DEVICE_STATUS_4, DEVICE_
STATUS_5, DEVICE_STATUS_6, DEVICE_STATUS_7 and DEVICE_STATUS_8 (index 51, 52,
53, 54, 55, 56, 57 and 58) in Physical Block.
l
FLXA21-PH
DEVICE_STATUS_1: Table 6.3
DEVICE_STATUS_2: Table 6.4
DEVICE_STATUS_3: Table 6.5
DEVICE_STATUS_4: Table 6.6
DEVICE_STATUS_5: Table 6.7
DEVICE_STATUS_6: Table 6.8
DEVICE_STATUS_7: Not used.
DEVICE_STATUS_8: Table 6.9