<12. OPERATION OF DO (Dissolved Oxygen)>
12-4
IM 12A01A02-01E
n
Configure sensor
On this screen, select the units and set the sensor sensitivity. If Polarograph is selected for the
sensor type, the polarograph application voltage can also be set. For details, see section 13.2.1.
For DO30G, set the sensor type to “DO30G”. The Available KOH parameter on the detail display
will be enabled.
n
Temperature settings
Select a suitable temperature element from among those displayed and set it up.
Celsius (ºC) or Fahrenheit (ºF) temperature scale can be selected.
For details, see section 13.2.2.
n
mA (output)
Select an appropriate process parameter from among those displayed and set it. If it is necessary
to improve the resolution, set it to a value suitable for the relevant process.
For details, see section 13.3.
12.3 Home display, Main display and Monitor
display
Pressing
changes the screen to the Main display shown in Figure 12.3 (or the Home
display shown in Figure 12.4).
If two sensors are connected, pressing
on the Main display changes the display to the
Home display shown in Figure 12.4.
If only one sensor is connected,
is grayed out and disabled on the Main display.
10.38
Tag:DO
25.0
4mA
20mA
Oxygen1
°C
mg/L
Figure 12.3
Example of main display
6.35
Tag:DO
24.9
4mA
20mA
Oxygen1
°C
mg/L
Tag:DO
25.0 °C
mg/L
10.38
Figure 12.4
Example of home display
On the Home display, pressing
of the 1st sensor (top) or 2nd sensor (bottom) causes the
display of the selected sensor to appear on the Main display.
8th Edition : Oct. 01, 2015-00