![YOKOGAWA FLXA202 User Manual Download Page 126](http://html.mh-extra.com/html/yokogawa/flxa202/flxa202_user-manual_911255126.webp)
<8. CALIBRATION OF SC (Conductivity)>
8-4
IM 12A01A02-01E
8th Edition : Oct. 01, 2015-00
8.8
Temporary output
If measurement cannot be made due to replacement of one of the sensors, assignment of mA
output can be temporarily changed from the currently setup sensor to the other sensor. This
feature is enabled only when two sensors are connected.
Press
to select Execute: Temporary output and then choose the process parameter to be
assigned on the Temporary output screen.
When process parameter assignment has been modified, the
indication at the upper left of the
screen changes to .
The feature is disabled the moment the screen returns to the Main display, and the setting
(process parameter in mA output setting) returns to the original condition.