4.9 Burst Setting and Manipulation
107
IM FG410-01EN
c) Auto burst setting screen
■
1st page: Screen for the basic parameters
These are the common setting items that are common to all the oscillation modes.
Oscillation start/stop phase [Phase]
This is the phase at which oscillation starts and stops.
■
2nd page: Screen for the auto burst settings
Mark wave number [Mark]
This is the wave number for oscillation. It can be set in 0.5-cycle units. Normally, it is
set in 1-cycle units.
Space wave number [Space]
This is the wave number for oscillation stop. It can be set in 0.5-cycle units. Normally, it
is set in 1-cycle units.
Stop level [StpLvl]
This is the signal level during oscillation stop. The level can be set to either off or on.
Normally, it is set to off.
Sync output [SyncOut]
This is the output signal from the sync/sub-output terminal. Waveform reference phase
synchronization and burst synchronization can be chosen from.
d) Waveforms for which auto burst is not possible
Auto burst is not possible for noise and DC.
e) Setting items required for auto burst
Set [Phase] (oscillation start/stop phase) on the 1st page of the setting screen.
Set [Mark[ (mark wave number) and [Space] (space wave number) on the 2nd page of the
setting screen.
Each wave number is normally set to an integer value.
[StpLvl] (stop level) on the 2nd page of the setting screen is normally set to [Off] (off).
Set the oscillation
mode to [Burst]
Press the NEXT
key to display
the 2nd page
Oscillation start/stop
phase setting
Indicates that the 1st page
is displayed
Mark wave number
Press the NEXT
key to display
the 1st page
Stop level
Indicates that the 2nd
page is displayed
Space wave number
Set burst mode to
[Auto]
Sync output