background image

User’s

Manual

FC800D

Non-reagent Type Free Available 

Chlorine Sensor Unit

Start-up and Safety Precautions

IM 12F05B10-01EN

IM 12F05B10-01EN

2nd Edition

Summary of Contents for FC800D

Page 1: ...User s Manual FC800D Non reagent Type Free Available Chlorine Sensor Unit Start up and Safety Precautions IM 12F05B10 01EN IM 12F05B10 01EN 2nd Edition...

Page 2: ...Attached to the product printed manual FLXA402T Liquid Analyzer for Turbidity and Chlorine Installation and Wiring IM 12A01G01 02EN Online manual FLXA402T Liquid Analyzer for Turbidity and Chlorine Op...

Page 3: ...y and service Yokogawa products and parts are guaranteed free from defects in workmanship and material under normal use and service for a period of typically 12 months from the date of shipment from t...

Page 4: ...0 u Control of Pollution Caused by the Product This is an explanation for the product based on Control of Pollution caused by Electronic Information Products in the People s Republic of China Pb Hg Cd...

Page 5: ...ging Repair work should be done at grounded workstations using grounded soldering irons and wrist straps to avoid electrostatic discharge CAUTION The Instrument is packed carefully with shock absorbin...

Page 6: ...safety symbols are used on the product as well as in this manual WARNING This symbol indicates that an operator must follow the instructions laid out in this manual in order to avoid the risks for th...

Page 7: ...allation Pollution level shows the level of existing the solid liquid and gas which break electrical isolation Category 2 is applied to normal room environment EMC EN 61326 1 Class A Table 2 For use i...

Page 8: ...insensitive version 0 1 to 0 3 mg L Free chlorine 0 5 mg L Free chlorine 0 10 mg L Type AB AD AG AJ General purpose for CE RCM China standard General purpose for CSA General purpose for KC General pur...

Page 9: ...ows for sufficient maintenance space with easy maintenance access allows for drainage For information on FLXA402T read the user s manual Install the FC800D s converter and sensor unit in pipes racks o...

Page 10: ...y Precautions 10 IM 12F05B10 01EN 2nd Edition June 01 2021 00 273 210 5 Unit mm Maintenance space Maintenance space Needle valve Flow cell Sample intle Rc1 4 2B pipe 60 5 Drain outlet Tube size 15 ID...

Page 11: ...C 31 C 33 NO S1 42 NC 41 C 43 NO S2 72 NC 71 C 73 NO S4 61 62 mA1 65 66 mA2 81 82 mA3 85 86 mA4 21 22 24 SHLD 63 87 88 SHLD 89 A 91 B 92 GND 93 SHLD 94 Termination 95 L 101 N 102 G 001 52 NC 51 C 53 N...

Page 12: ...stance 100 ohm or less 2 Connection cable is normally 1 m in length It is available with L02 L03 L05 L10 or L20 depending on the length you need 3 In case of selecting TB Terminal box for connecting a...

Page 13: ...les of power supply by using stanchion cable rack conduit or vinyl band Unplugged cables are dangerous and may cause an electric shock CAUTION When you open the front cover on the detector and Relay B...

Page 14: ...2T Installation and Wiring IM 12A01G01 02EN n Wiring the power supply The cable size of terminals is 0 13 to 2 5 mm2 in cross sectional area or AWG 26 to 14 Attached cable gland is for cables finished...

Page 15: ...e Start a regular measurement Refer to IM 12F05B10 02EN for details 4 Maintenance n Periodic maintenance When you must open the front cover and or glands make sure that the seals are clean and correct...

Page 16: ...ne 01 2021 00 Revision record June 2021 2nd Edition Revised compliant standards p 4 7 Added a code to Model and Suffix Codes p 8 Mar 2021 1st Edition Newly released Yokogawa Electric Corporation 2 9 3...

Reviews: