
<8. Detailed Data Setting>
8-7
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
In a semi-automatic calibration, “during calibration” is the time, starting when a calibration
instruction is executed with an infrared switch or a contact input, to make a series of
calibrations, until the preset output stabilization time elapses.
In an automatic calibration, “during calibration” is the time, starting when automatic
calibration is carried out at the calibration start time, until the preset output stabilization time
elapses.
(4) “Error” appears when Error 1 to Error 4 are being issued
8.2.2
Preference Order of Output Hold Value
The output hold value takes the following preference order:
During error occurrence
During calibration
During maintenance
During warm-up
Preference order (high)
8.3.2E.siki
For example, if the output current is set to “4 mA” during maintenance, and “without hold” output
during calibration is preset, the output is held at 4 mA in the maintenance display. However, the
output hold is released at the time of starting the calibration, and the output will be held again at 4
mA after completing the calibration and when the output stabilization time elapses.
8.2.3
Output Hold Setting
Table 8.5 lists parameter codes with set values for individual set items.
Table 8.5
Parameter Codes for Output Holding
Set items
Parameter code
Set value
During warm-up
C04
0 4 mA
1 20 mA
2 Holds Set value
During maintenance
C05
0 Without hold feature
1 Last measured value.
2 Holds set values.
During calibration
C06
0 Without hold feature
1 Last measured value.
2 Holds set values.
During error occurrence
C07
0 Without hold feature
1 Last measured value.
2 Holds set values.
Note: “C07” is not displayed when option code “/C2” or “/C3” (NAMUR NE 43 compliant) is specified.
8.2.4
Default Values
When the analyzer is delivered, or if data are initialized, output holding is by default as shown in
Table 8.6.
Table 8.6
Output Hold Default Values
Status
Output hold (min. and max. values)
Preset value
During warm-up
4 mA
4 mA
Under maintenance
Holds output at value just before maintenance started.
4 mA
Under calibration or blow-back Holds output at value just before starting calibration
4 mA
On Error occurrence
Holds output at a preset value.
3.4 mA
Summary of Contents for EXAxt ZR402G
Page 9: ...Blank Page...
Page 51: ...Blank Page...
Page 151: ...Blank Page...
Page 165: ......
Page 167: ......