<6. Components>
6-2
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
6.1.2
High Temperature Detector (ZR22G-015)
Sample gas outlet
When a sample gas pressure is
negative, connect the ejector assembly.
When the sample gas is high
temperature and high pressure, and does
not fall below 700°C, connect a pressure
control valve (e.g. a needle valve).
(Refer to Section 3.2.1.)
Separate type
High Temperature Detector (ZR22G-015)
When the temperature of the sample gas is
between 700°C and 1400°C, mount this
detector with a ZO21P-H probe adapter.
F6.2E.ai
Flange
Selectable from JIS, ANSI,
JPI or DIN
High Temperature Probe Adapter (ZO21P-H)
The probe is made of either SUS310S
or silicon carbide (SiC).
Its length is 0.5, 0.6, 0.7, 0.8, 0.9, 1.0 or 1.5m.
When using an SiC probe, mount it vertically
downward.
Figure 6.2
High Temperature Detector