<5. Wiring>
5-9
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
5.2.6
Wiring for Analog Output
This wiring is for transmitting 4 to 20 mA DC output signals to a device, e.g. recorder.
Maintain the load resistance including the wiring resistance at 550 Ω or less.
Shielded cable
Receiver 1
Receiver 2
ZR402G
Converter
1
2
1
2
AO-1(+)
AO-1(-)
AO-2(+)
AO-2(-)
FG
F5-9E.ai
Figure 5.8
Wiring for analog output
Cable Specifications
For this wiring, use a 2-core or a 4-core shielded cable.
Wiring Procedure
(1) M4 screws are used for the terminals of the converter. Each wire in the cable should be
terminated corresponding to crimp-on terminals. Ensure that the cable shield is connected
to the FG terminal of the converter.
(2) Be sure to connect “+” and “-” polarities correctly.
5.2.7
Contact Output Wiring
Contact outputs 1 to 3 can be freely assigned to “low limit alarm”, “high limit alarm”, etc. user
selectable, but the assignment of contact output 4 is fixed (“error output”). And the action (contact
closed on error output) also cannot be changed.
When using these contact outputs, install the wiring as follows:
#1 Output
#2 Output
#3 Output
#4 Output
DO-1
DO-1
DO-2
DO-2
DO-3
DO-3
Terminal box
Annunciator or the like
F5-11E.ai
ZR402G
Converter
DO-4
DO-4
Figure 5.9
Contact output wiring
Cable Specifications
Number of wire in cable varies depending on the number of contact used.
Summary of Contents for EXAxt ZR22G
Page 15: ...Blank Page...
Page 51: ...Blank Page...
Page 77: ...Blank Page...
Page 137: ...Blank Page...
Page 171: ...Blank Page...
Page 195: ...Blank Page...
Page 197: ......
Page 199: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 207: ......
Page 211: ...Blank Page...