<5. Wiring>
5-5
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
5.2 Wiring
5.2.1
Connection to Converter
To connect the wiring to the converter, proceed as follows:
(1) M4 screws are used for the terminals of the converter. Each cable should be terminated in
the corresponding size crimp-on terminals.
(2) When a rubber insulated glass braided wire is used for wiring to the detector, use a terminal
box. For wiring between the terminal box and the converter, basically use a cable that
withstand temperatures of at least 80°C.
NOTE
The above is to prevent moisture or corrosive gas from entering the converter.
Where the ambient environment of the detector and the converter is well-maintained, it is
permissible to connect the wiring from the detector directly to the converter with protection by
conduits.
WARNING
This wiring is to carry power for the heater. Be careful to wire the correct terminals, and be
careful not to ground or short circuit terminals when wiring, as otherwise the instrument may be
damaged.
5.2.2
Connection to Detector
When connecting the cable to the detector, proceed as follows:
(1) Mount cable glands or conduits of the specified thread size to the wiring connections of the
detector.
The detector may need to be removed in future for maintenance, so be sure to allow
sufficient cable length.
(2) If the ambient temperature at the location of wire installation is 75 to 150
°
C, be sure to use a
flexible metallic conduit for the wire. If a non-shielded “ 600 V silicon rubber insulated glass
braided wire” is used, keep the wire away from noise sources to avoid noise interference.
(3) The size of the terminal screw threads is M3.5. Each cable should be terminated in the
corresponding size crimp-on terminals contact (*1) respectively.
*1 If the ambient temperature at the detector installation site exceeds 60°C, use a “ bare crimp-on terminal”.
l
Notice when closing the cover of the detector
NOTE
• Before opening the detector cover, loosen the lock screw. If the screw is not loosened first,
the screw will damage the cover, and the terminal box will require replacement.
When opening and closing the cover, remove any sand particles or dust to avoid gouging
the thread.
• After screwing the cover in the detector body, secure it with the lock screw.
Summary of Contents for EXAxt ZR22G
Page 15: ...Blank Page...
Page 51: ...Blank Page...
Page 77: ...Blank Page...
Page 137: ...Blank Page...
Page 171: ...Blank Page...
Page 195: ...Blank Page...
Page 197: ......
Page 199: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 207: ......
Page 211: ...Blank Page...