<5. Wiring>
5-2
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
Table 5.1
Cable specifications
Terminal name of converter
Name
Need for shields Number of cores
CELL+, CELL-
HTR TC+, HTR TC-
CJ+, CJ-
Detector signal
О
6
HEATER
Detector heater
2
L, N
Power supply
2 or 3 *
AO-1+, AO-1-, AO-2+, AO-2-
Analog output
О
2 or 4
DO-1, DO-2, DO-3, DO-4
Contact output
2 to 8
AC-Z, AC-S, AC-C
Automatic Calibration unit
3
DI-1, DI-2, DI-C
Contact input
3
AI+, AI-
Temperature input
О
2
Note *: When the case is used for protective grounding, use a 2-core cable.
WARNING
Cables that withstand temperatures of at least 80°C should be used for wiring.
CAUTION
• Select suitable cable O.D. to match the cable gland size.
• Protective grounding should be connected in ways equivalent to JIS D style (Class 3)
grounding (the grounding resistance is 100 Ω or less).
• Special consideration of cable length should be taken for the HART communication, For the
detail, refer to Section 1.1.2 of the IM 11M12A01-51E “Communication Line Requirements”.
5.1.1
Terminals for the External Wiring in the Converter
Open the front door and remove the terminal covering plate to gain access to the converter
external wiring terminals (see Figure 5.1).
CAUTION
After wiring necessary cable to the converter terminals, be sure to fix the terminal covering plate
with two screws again.
Summary of Contents for EXAxt ZR22G
Page 15: ...Blank Page...
Page 51: ...Blank Page...
Page 77: ...Blank Page...
Page 137: ...Blank Page...
Page 171: ...Blank Page...
Page 195: ...Blank Page...
Page 197: ......
Page 199: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 207: ......
Page 211: ...Blank Page...