<11. Inspection and Maintenance>
11-7
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
Part No.
Description
(7)
K9470BJ
Metal O-ring
(14)
K9470ZS
O-ring with grease
(21)
(22)
K9470ZP
Two pairs of O-rings with grease
11.1.6
Cleaning the High Temperature Probe Adapter
CAUTION
Do not subject the probe of the High Temperature Probe Adapter (ZO21P-H-A) to shock. This
probe uses silicon carbide (SiC) which may become damaged if it is subjected to a strong shock
or thermal shock.
The high temperature detector is structured so that the gas to be measured is directed toward
the detector with the high temperature probe adapter. Therefore, if the probe or the sample gas
outlet clogs, a precise measurement is no longer possible because of no gas flow. If you use the
high temperature detector, you have to inspect it periodically and, if any part of it is significantly
clogged with dust, clean it.
Dust found sticking to the probe should be blown off. If any dust still remains after the blowing,
clean it with a metal rod, etc., inserted. In addition, if dust is found on the auxiliary ejector or
needle valve (choke) at the sample gas outlet, remove these parts from the high temperature
probe adapter and then clean them. To remove dust, blow air on them or rinse them with water.
11.1.7
Stopping and Re-starting Operation
<Stopping Operation>
When operation is stopped, take care of the followings so that the sensor of the detector cannot
become unused.
CAUTION
When operating an instrument such as boiler or industrial furnace is stopped with the zirconia
oxygen analyzer operation, moisture can condensate on the sensor portion and dusts may stick
to it.
If operation is restarted in this condition, the sensor which is heated up to 750°C firmly fixes the
dusts on itself. Consequently, the dusts can make the sensor performance much lower. If a large
amount of water is condensed, the sensor can be broken and will never be used.
To prevent the above nonconformity, take the following action when stopping operation.
(1)
If possible, keep on supplying the power to converter and flowing reference gas to the sensor.
If this is impossible to do the above, remove the detector.
(2) If unavoidably impossible to supply the power and removing the detector, keep on the
following air at 600 ml/min into the calibration gas pipe.
<Restarting Operation>
When restarting operation, be sure to flow air, for 5-10 minutes, at 600 ml/min into the calibration
gas pipe before supplying power to the converter.
Summary of Contents for EXAxt ZR22G
Page 15: ...Blank Page...
Page 51: ...Blank Page...
Page 77: ...Blank Page...
Page 137: ...Blank Page...
Page 171: ...Blank Page...
Page 195: ...Blank Page...
Page 197: ......
Page 199: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 207: ......
Page 211: ...Blank Page...