<10. Other Functions>
10-9
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
Analog output remains held
(with HOLD feature specified)
ON-time period of 1 to 11 seconds for the contact input
Blow back start
(with contact input or
touchpanel operation)
Blow back time
Contact output
(blow back switch)
ON
(for 10
seconds)
ON
(for 10
seconds)
OFF
(for 10 seconds)
Analog output
Hold time
ON
(for 10
seconds)
F10.9E.ai
Figure 10.9
Operation of Blow back
10.4.3
Setting Output Hold Time and Blow back Time
If the blow back mode is in “No function”, the output “Hold time” and “Blow back time” are not
displayed. If you select “Hold time”, the numeric-data entry display appears. Enter the desired
“Hold time” (output stabilization time) from 00 minutes, 00 seconds to 60 minutes, 59 seconds.
When you select “Blow back time”, the numeric-data entry display appears. Enter the desired
“Blow back time” from 00 minutes, 00 seconds to 60 minutes, 59 seconds.
10.4.4
Setting Interval, Start Date, and Start Time
The “Interval” is the time to execute blow back. Display the numeric-data entry panel display to
set the desired interval from 000 days, 00 hours to 255 days, 59 hours.
For the “Start date” and “Start time”, set the date when the blow back is first executed and the
time when to start the blow back, respectively. If you want to execute the first blow back, for
example, at 4:00 p.m. on March 25, 2001, enter 25/ 03/ 01 for the Start date and 16:00 for the
Start time.
Mode: Auto
Hold time: 1 0 min 0 0 s
Blow back time:
1 0 min 0 0 s
Interval: 3 0 d 0 0 h
Start date: 0 1 / 0 1 / 0 0
Start time: 0 0 : 0 0
Enter
Blow back setup
F10.10E.ai
Figure 10.10
Blow back Setup Display
In the Blow back setup display shown in Figure 10.10, if you choose “Mode: No function” or
“Semi_Auto”, the Interval, Start date, and Start time for these are not displayed.
NOTE
• If the blow back is executed with an input contact, it must be preset in the Input contact
settings (for more details, see Section 8.5, earlier in this manual).
• In Section 8.4, Output Contact Setup, earlier in this manual, set the contact used as the
blow back switch beforehand.
• Do not set any other function for the contact used as the blow back switch.
Otherwise, blow back may be activated when the contact is closed by any other function.
Summary of Contents for EXAxt ZR22G
Page 15: ...Blank Page...
Page 51: ...Blank Page...
Page 77: ...Blank Page...
Page 137: ...Blank Page...
Page 171: ...Blank Page...
Page 195: ...Blank Page...
Page 197: ......
Page 199: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 207: ......
Page 211: ...Blank Page...