<7. Startup>
7-6
IM 11M12A01-02E
(3) To set the minimum humidity at 4 mA, choose the Min. humidity conc. To set the maximum
humidity at 20 mA, choose the Max. humidity conc.
(4)
To set 50% H
2
O, type in 050 and press the [Enter] key.
(5)
Set “mA-output2” in the same manner as in the appropriate steps above.
NOTE
Analog output range settings should be limited. For more details, consult Section 8.1, “Current
Output setting,” later in this manual.
7.9
Setting Display Item
7.9.1
Oxygen Analyzer - Setting Display Item
This section briefly describes the display item settings shown in Figure 7.13, “Basic Panel
Display.”
F7.11E.ai
Tag:
21.0
%O
2
17.43mA -Output1
17.43mA -Output2
Primary value
Secondary value
Tertiary value
Figure 7.13
Basic Panel Display
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select “Maintenance” in the Execution/Setup display.
(2) Select the “Display setup” from the Maintenance panel display (Figure 7.14).
The Display setup display (Figure 7.15) then appears.
(3) In the above Display setup display, select the “Display item”. The Display item display
(Figure 7.16) then appears. From this display, select the “Primary value” and press the
[Enter] key to display the “Display item” selection display (Figure 7.17).
(4) Select the Secondary and Tertiary values in the same manner as in the steps above.
(5) Consult Table 7.1, Display Items, enabling the selection of display items in individual display
areas.
16th Edition : Mar. 19, 2020-00
Summary of Contents for EXAxt ZR22G
Page 15: ...Blank Page...
Page 51: ...Blank Page...
Page 77: ...Blank Page...
Page 137: ...Blank Page...
Page 171: ...Blank Page...
Page 195: ...Blank Page...
Page 197: ......
Page 199: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 207: ......
Page 211: ...Blank Page...