<7. MAINTENANCE>
7-7
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
7.6.2 Zero Calibration by Turning Off Light Source
If the turbidity to be measured is 200 mg/L or more, zero calibration can be performed by turning
OFF the light source.
The zero-point difference between this method and the "zero water" method is about 1 mg/L.
(1) Select FUNCTION "2" in the <MAINT.> mode.
=> "Turbidity value" is displayed.
(2) The lamp goes out when the [ENT] key is pressed once.
(3) Wait one minute or more until the reading becomes stable, and then press the [>] and [ENT]
keys in this order, Now calibration is over.
(4) Finished.