<7. MAINTENANCE>
7-4
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
7.5 Lamp Replacement
The lamp's service life is more than year. However, it is recommended that it be replaced with a
new one after a period of one year as preventive maintenance.
This section describes replacement of the lamp.
(1) Turn off the power switch.
(2) Remove the four C screws in the upper part of the detector and remove the cover (see
Figure 7.2).
(3) Disconnect terminals P1 and P2 by removing them from the terminal board and remove the
two B screws, then remove the lamp holder.
(4) Loosen the clamp that locks the lamp.
(5) Replace the lamp. In doing this, confirm that the boss of the new lamp is placed in the
position shown in Figure 7.2.
Cover
Four C screws
Clamp
Two B screws
Lamp holder
Lamp boss
Side view of detector
Front view of detector
Terminals P1 and P2
Terminal AG
Terminal S1
P1 P2
F0703.ai
Figure 7.2 Lamp replacement
(6) Temporarily set the clamp lightly in place for later adjustment.
(7) Using a cotton swab soaked in alcohol or relevant cleaning solution, clean the backside of
the lens in the detector.
(8) Tighten terminals P1 and P2 to the terminal board and also the two B screws.
(9) Turn on the power. Wait about 5 minutes or more until the brightness of the lamp has
stabilized.
(10) Carry out lamp control reference value calibration.
Select FUNCTION "C" in the <MAINT.> mode.