IM 01C25H01-01E
7-4
7. MAINTENANCE
7.4.2 Replacing the CPU Board Assembly
This subsection describes the procedure for replacing
the CPU assembly. (See figure 7.3)
Removing the CPU Assembly
1) Remove the cover. If an integral indicator is
mounted, refer to subsection 7.4.1 and remove the
indicator.
2) Turn the zero-adjustment screw to the position
(where the screw head slot is horizontal) as shown
in figure 7.3.
3) Disconnect the output terminal cable (cable with
brown connector at the end). When doing this,
lightly press the side of the CPU assembly connec-
tor and pull the cable connector to disengage.
4) Use a socket driver (width across flats, 5.5mm) to
loosen the two bosses.
5) Carefully pull the CPU assembly straight forward to
remove it.
6) Disconnect the flat cable (cable with white connec-
tor at the end) that connects the CPU assembly and
the capsule.
NOTE
Be careful not to apply excessive force to the
CPU assembly when removing it.
Mounting the CPU Assembly
1) Connect the flat cable (with white connector)
between the CPU assembly and the capsule.
2) Connect the output terminal cable (with brown
connector).
NOTE