<10. Parameter Summary>
10-1
IM 01C27F01-01EN
10. Parameter Summary
The setting of differential pressure (Al2 block) is invalid in this instrument. The output value of Al2 object and
the output status are “0.0”(zero) and “Uncertain” respectively.
Various diagnostics of the Al2 object and the static pressure is not supported in this instrument.
Table 10.1
Parameter
Object
ID
Attribute
ID
Label
Description
Default
value
Handling
1.
UAPMO
block
1
Version Revision
Indicates the application revision of EJX This revision
when the application software is downloaded.
---
R
10
Static Revision
Indicates the revision level of the fixed parameters of
UAP Used, for example, to check whether parameters
have been change.
0
R
64
Identification
Number
Indicates the vender ID, model ID, and revision of the
device.
---
R
65
CTS Version
Indicates the version of the communication stack test
system (CTS).
0
R
66
ITS Version
Indicates the version of the inter operability test system
(ITS).
0
R
67
Diagnostic Status
Indicates the diagnostic results of the device based on
the NAMUR NE107
*1
model.
Setting Enable diagnostic status configuration in UAP
Option to Enable allows turning OFF and ON the
display of the diagnostic results for each summary,
and changing Categorize For Categorize at the time of
shipment, refer to tables 8.3 and 10.2.
---
R
68
UAP Option
Allows setting the Diagnostic Status and write protection
of UAP.
Software write protect
1: On, 0: Off
Enable hardware write protect
1: Enable, 0: Disable (default)
Enable diagnostic status configuration
1: Enable, 0: Disable (default)
The following table shows the relationship between the
hardware write protection and software write protection.
Enable
hardware
write
protect
Hardware
write
protect
Software
write
protect
Write
protect
Disable
Off or On
Off
No
(Writable)
Disable
Off or On
On
Yes
(Unwritable)
Enable
Off
Off or On
No
(Writable)
Enable
On
Off or On
Yes
(Unwritable)
0
W
69
Diagnostic
Switch
Allows setting On/Off for each summary of Diagnostic
Status when Enable diagnostic status configuration in
UAP Option is set to Enable.
On
W
70
Diagnostic
Configuration
Allows Categorize for each summary of Diagnostic
Status when Enable diagnostic status configuration in
UAP Option is set to Enable.
Refer to Table
10.2.
W
102
Diagnostic Status
Detail[2]
Detailed information on Diagnostic Status. For
Categorize at the time of shipment, refer to Table 10.2.
Refer to Table
10.2.
R
103
Energy Left
Indicates the number of days of remaining battery life
assuming ambient temperature condition as 23 degrees
Celsius.
The unit is day.
---
R