![YOKOGAWA DP harp EJX210B User Manual Download Page 92](http://html1.mh-extra.com/html/yokogawa/dp-harp-ejx210b/dp-harp-ejx210b_user-manual_3522930092.webp)
<
11. General Specifications
>
11-10
IM 01C27C01-01EN
11.3 Optional Specifications
OPTIONAL SPECIFICATIONS (For Explosion Protected type)
Item
Description
Code
Factory Mutual (FM)
FM Intrinsically safe Approval
FS17
ATEX
ATEX Intrinsically safe Approval
KS27
Canadian Standards Association (CSA)
CSA Intrinsically safe Approval
CS17
IECEx
IECEx Intrinsically safe Approval
SS27
OPTIONAL SPECIFICATIONS (Others)
Item
Description
Code
Painting Color change
Amplifier cover only
P
Coating change Anti-corrosion coating
*1
X2
Oil-prohibited use
Degrease cleansing treatment
K1
Oil-prohibited use with
dehydrating treatment
Degrease cleansing and dehydrating treatment
K5
Calibration units
*2
P calibration (psi unit)
(See table for Span and
Range Limits.)
D1
bar calibration (bar unit)
D3
M calibration (kgf/cm
2
unit)
D4
Teflon film
*3 *4
Diaphragm protection from sticky process fluid by FEP Teflon film attached with
fluorinated oil.
Operation range: 20 to 150°C, 0 to 2 MPa (Not applicable for vacuum service).
TF1
Gold-plated diaphragm
*5
Inside of isolating diaphragms (fill fluid side) are gold plated, effective for
hydrogen permeation.
A1
Material
certificate
For Flush type
High Pressure side: Process flange, Block
*6
Low Pressure side: Cover flange
M0W
High Pressure side: Process flange, Block
*7
Low Pressure side: Cover flange, Process connector
M1W
High Pressure side: Process flange, Block, Ring
*6 *8
Low Pressure side: Cover flange
M3W
High Pressure side: Process flange, Block, Ring
*7 *8
Low Pressure side: Cover flange, Process connector
M4W
For Extended
type
High Pressure side: Process flange, Block, Pipe, Base
*6
Low Pressure side: Cover flange
M0E
High Pressure side: Process flange, Block, Pipe, Base
*7
Low Pressure side: Cover flange, Process connector
M1E
Pressure test/Leak test
Certificate
*9*10
(Flange rating)
(Test pressure)
JIS 10K
2 MPa (290 psi)
Nitrogen (N
2
) Gas
*13
Retention time: one minute
T51
JIS 20K
5 MPa (720 psi)
T54
ANSI/JPI Class 150
3 MPa (430 psi)
T52
ANSI/JPI Class 300
8 MPa (1160 psi)
*11
T56
ANSI/JPI Class 300
7 MPa (1000 psi)
*12
T55
*1:
Not applicable with color change option.
*2:
The unit of MWP (Max. working pressure) on the name plate of a housing is the same unit as specified by option code
D1
,
D3
, and
D4
.
*3:
Applicable for flush type (process connection style code
W
.)
*4:
Applicable for flushing connection ring code
0
.
*5:
Applicable for wetted parts material code
SW
,
SE
,
WW
,
WE
, and
HW
.Consult Yokogawa in case gold-plated diaphragm is required
for low pressure side.
*6:
Applicable for Low Pressure Side Process connection code
0
and
5
.
*7:
Applicable for Low Pressure Side Process connection code
1
,
2
,
3
, and
4
.
*8:
Applicable for flushing connection ring code
A
,
B
,
C
, and
D
.
*9:
The unit on the certificate is always MPa regardless of selection of option code
D1
,
D3
, or
D4
.
*10: A flushing connection ring will not be applied when conducting the pressure test or leak test.
*11: Applicable for flush type (process connection style code
W
.)
*12: Applicable for extended type (process connection style code
E
.)
*13: Pure nitrogen gas is used for oil-prohibited use (option code
K1
and
K5
.)