<11. General Specifications>
11-1
IM 01C27C01-01EN
11. General Specifications
11.1 Standard Specifications
Communication protocol:
ISA100.11a protocol
Data rate:
250 kbps
Frequency:
2400 - 2483.5 MHz license free ISM band
Radio security:
AES 128 bit codified
RF Transmitter power:
Max. 11.6 dBm (fixed)
Antenna:
+2 dBi Omni directional monopole type
Span and range limits:
Measurement
Span/Range
kPa
inH
2
O(/D1)
mbar(/D3) mmH
2
O(/D4)
M
Span
1 to 100
4 to 400
10 to 1000
100 to 10000
Range -100 to 100 -400 to 400 -1000 to 1000
-10000 to
10000
H
Span
5 to 500
20 to 2000
50 to 5000
0.05 to
5 kgf/cm
2
Range -500 to 500 -2000 to 2000 -5000 to 5000
-5 to
5 kgf/cm
2
Performance specifications:
Refer to GS 01C27C01-01EN.
Update Time
Measurement
mode
Differential
pressure
Pressure
Continuous
100 ms
100 ms
Intermittent
0.5 to 3600 s
selectable
0.5 to 3600 s
selectable
The transmitter shifts to the countinuous mode
when the update time is set to 0.5 second.
Power Supply Specifications
Battery:
Use the dedicated battery pack.
Rated voltage: 7.2 V
Rated capacity: 19 Ah
Output:
ISA100.11a protocol
Output mode:
linear or square root
Ambient Temperature Limits:
-40 to 85°C (-40 to 185°F)
-30 to 80°C (-22 to 176°F) LCD visible range
(Note: The ambient temperature limits must be
within the fill fluid operating temperature range,
see table 11.1.)
Process Temperature Limits:
High pressure side: See table 11.1.
Low pressure side: 40 to 120°C (-40 to 248°F)
Ambient Humidity Limits:
0 to 100% RH
Working Pressure Limits
See table 11.1.
For atmospheric pressure or below, see figure
11.1.
2.7 (0.38)
100 (14.5)
1 (0.14)
10 (1.4)
–50
(–58)
0
(32)
50
(122)
100
(212)
150
(302)
200
(392)
250
(482)
300
(572)
F1101.ai
Process temperature range
Flange max.
working
pressure
Atmospheric
pressure
Transmitter ambient
temperature range
Process Temperature °C (°F)
Working
pressure
kPa abs
(psi abs)
Figure 11.1 Working Pressure and Process
Temperature