Optical Dissolved Oxygen, model DO71/DO72

 

Information

 

MODBUS Communication

 

 

TI 12J06J01-00EN-P 

 

Yokogawa Process Analyzers Europe B.V. 

TI 12J06J01-00EN-P, 1st Edition 

 

             Euroweg 2, 3825 HD Amersfoort, The Netherlands 

Copyright © 2022 

 

 

 

 

 

 

Summary of Contents for DO71

Page 1: ...l Dissolved Oxygen model DO71 DO72 Information MODBUS Communication TI 12J06J01 00EN P Yokogawa Process Analyzers Europe B V TI 12J06J01 00EN P 1st Edition Euroweg 2 3825 HD Amersfoort The Netherlands...

Page 2: ...2 All Rights Reserved Copyright 2022 Yokogawa Process Analyzers Europe B V TI 12J06J01 00EN P This page is empty...

Page 3: ...rs Europe B V TI 12J06J01 00EN P Contents Introduction 4 1 General 5 1 1 MODBUS setup 5 1 2 Function code Exception code 5 1 3 MODBUS communication 6 1 4 Connection 6 2 MODBUS map 8 2 1 Hold Register...

Page 4: ...xygen sensors model DO71 and model DO72 IM 12J06J01 00EN P for details of the parameters and connections Latest revision of the User s Manual can be downloaded from the website https www yokogawa com...

Page 5: ...ment Slave address 1 Serial Profile 19200 bps 8 data bits No parity 2 stop bits NOTE Slave address and Serial Profile are configurable as explained in this document In case of using multiple Optical D...

Page 6: ...The Optical Dissolved Oxygen sensor s HR HOLD Register response to a HOST HR Read request is within 100ms The timing in between a response and a new request must be at least 10ms Figure 1 3 Timing di...

Page 7: ...tion 12 yellow RS485 Data B 11 green RS484 Data A 13 white Power GND 10 brown Power VDC 14 transparent Cable shield Probe body Table 1 4 2 DO72 connection definition Terminal pin no WU10 V D xx Wire c...

Page 8: ...CDAB ASCII character set using one byte per character multiple registration possible little endian byte swap CDAB Access load import to multiple data types such as Float Uint32 ASCII etc must be impl...

Page 9: ...de 16 bytes 8 addresses 2089 Oxygen unit Uint32 R W 0x 0000 0010 vol O2 0x 0000 0020 air saturation 0x 0000 0040 ppb g L 2 0x 0000 0080 ppm mg L 2 0x 0400 0000 mol L 2 0x 0800 0000 DO saturation 0x 20...

Page 10: ...y default this is the Persistent Pressure Comp value register HR3153 3153 Persistent Pressure Comp Value Float R W Default setting for compensation pressure value in hPa mbar Remark 1 Do not use this...

Page 11: ...d 0x 0000 0007 115200 baud Remark 1 Setting is effective immediately D 19200 L 4800 H 115200 4103 Reserved Uint32 R 4105 Reserved Uint32 R 4107 Parity Uint32 R W 0x 0000 0000 Even parity 1 Stop bit 0x...

Page 12: ...ensor was out of the defined measurement temperature range see registers HR4611 and HR4613 D 0 4679 Hours out of Operating Temperature Range Float R Accumulated hours in which the sensor was out of th...

Page 13: ...19999 H 0 19999 4915 Reserved Uint32 R D NaN 4917 Sensor Constant dPhi2 Float R W Sensor constant dPhi2 Remark 1 The default value is printed on the Quality Inspection Certificate Q I C L 0 19999 H 0...

Page 14: ...alues or the manually set temperature value will be written in the registers for Cal0 register HR5521 and T0 register HR5523 5181 Low Calibration Point Timestamp Uint32 R The timestamp the last low po...

Page 15: ...Oxygen concentration unit defined in HR5535 at the high oxygen calibration point L 0 001 H 299 999 5529 Cal 2nd Float R W Calibration value of Phase shift at the high oxygen calibration point L 10 01...

Page 16: ...reboot this timestamp is reset to 0 and must be set again to a value representing local time 2 Register is automatically incremented every second 3 Timestamp will be saved to the applicable calibratio...

Page 17: ...000008 Signal channel overflow Signal amplitude too high Decrease signal current 4 0x00000010 Reserved 5 0x00000020 Reserved 6 0x00000040 No sensor calculation Amplitude too low Signal amplitude is to...

Page 18: ...ion 0 0x00000000 Reserved 1 0x00000002 Reserved 2 0x00000004 Temperature value too low while calibrating the low point 3 0x00000008 Temperature value too high while calibrating the low point 4 0x00000...

Page 19: ...MODBUS Communication Revision History 19 All Rights Reserved Copyright 2022 Yokogawa Process Analyzers Europe B V TI 12J06J01 00EN P 5 Revision History October 2022 1st Edition...