4-2
IM DLM5058-03EN
4.2 Trigger Output (TRIGGER OUT)
CAUTION
Do not short the TRIGGER OUT terminal or apply external voltage to it. If you do, the
instrument may malfunction.
French
ATTENTION
Ne pas mettre la borne TRIGGER OUT en court-circuit et ne pas y appliquer une tension
externe. Si vous le faites, l’instrument risque de ne pas fonctionner correctement.
Trigger Output Terminal (TRIGGER OUT)
A 3.3 V CMOS level signal is output when the instrument triggers. The signal level is normally high
but goes low when a trigger occurs.
Item
Specifications
Connector type
BNC
Output level
3.3 V CMOS
Output impedance
Approx. 50 Ω
Output logic
Negative logic ( ) and positive logic ( ) switchable
Output delay time
600 ns or less*
Output hold time
For negative logic, the low level is 500 ns min. and the high level is 50 ns min.
For positive logic, the high level is 500 ns min. and the low level is 50 ns min.
* When deskewing is not used
Output Timing
Trigger occurrence
600 ns max
H
L
Trigger output
(negative logic)
Trigger signal
H
L
(positive logic)