5-174
IM DLM4038-17EN
:SERialbus<x>:CXPI:SEARch:SELect
Function Sets which detected point to display in the CXPI
bus signal search zoom window and queries the
zoom position of the detected point.
Syntax
:SERialbus<x>:CXPI:SEARch:
SELect {<NRf>|MAXimum}
:SERialbus<x>:CXPI:SEARch:SELect?
<x> = 1 to 4
<NRf> = 0 to 49999
Example
:SERIALBUS1:CXPI:SEARCH:SELECT 1
:SERIALBUS1:CXPI:SEARCH:SELECT?
-> :SERIALBUS1:CXPI:SEARCH:
SELECT 1.50000000
Description If there are no detected points, the DLM4000
returns “NAN.”
:SERialbus<x>:CXPI:SEARch:SELe
ct? MAXimum
Function Queries the number of detected points in the
CXPI bus signal search.
Syntax
:SERialbus<x>:CXPI:SEARch:
SELect? {MAXimum}
<x> = 1 to 4
Example
:SERIALBUS1:CXPI:SEARCH:
SELECT? MAXIMUM
-> :SERIALBUS1:CXPI:SEARCH:
SELECT 100
Description If there are no detected points, the DLM4000
returns “NAN.”
:SERialbus<x>:CXPI:SEARch:SETup?
Function Queries all CXPI bus signal search condition
settings.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup?
<x> = 1 to 4
:SERialbus<x>:CXPI:SEARch:SETup:ERR
or?
Function Queries all CXPI bus signal search error settings.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor?
<x> = 1 to 4
:SERialbus<x>:CXPI:SEARch:SETup:ERRo
r:CLOCk
Function Sets or queries the CXPI bus signal search clock
error setting.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:CLOCk {<Boolean>}
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:CLOCk?
<x> = 1 to 4
Example
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
CLOCK ON
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
CLOCK?
-> :SERIALBUS1:CXPI:SEARCH:SETUP:
ERROR:CLOCK 1
:SERialbus<x>:CXPI:SEARch:SETup:ERRo
r:COUNter
Function Sets or queries the CXPI bus signal search
counter error setting.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:COUNter {<Boolean>}
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:COUNter?
<x> = 1 to 4
Example
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
COUNTER ON
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
COUNTER?
-> :SERIALBUS1:CXPI:SEARCH:SETUP:
ERROR:COUNTER 1
:SERialbus<x>:CXPI:SEARch:SETup:ERRo
r:CRC
Function Sets or queries the CXPI bus signal search CRC
error setting.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:CRC {<Boolean>}
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:CRC?
<x> = 1 to 4
Example
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
CRC ON
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
CRC?
-> :SERIALBUS1:CXPI:SEARCH:SETUP:
ERROR:CRC 1