22 Other Features
Auto Setup (AUTO SETUP)
The auto setup feature automatically sets the SCALE, TIME/DIV, trigger level, and other settings to the most
suitable values for the input signals.
This feature is useful when you are not sure what type of signal will be applied to the instrument. The auto setup
feature will not work properly on some input signals.
Center Position after the Execution of Auto Setup
The center position after you execute auto setup will be 0 V.
Source Channels
Auto setup is performed on all channels except the logic channels.
Waveforms Displayed before the Execution of Auto Setup
Waveforms that were displayed before you execute auto setup will be cleared.
Signals That Auto Setup Can Be Applied To
You can use auto setup for the following types of input signals.
•Simple, repeating signals whose frequency is 50 Hz or higher
•Absolute input voltage: Signals whose maximum value is at least approx. 20 mV (at 1:1 setting)
• The auto setup feature may not work properly for signals that include a DC component or high-frequency
components.
• The auto setup menu for analyzing serial bus signals is located in the setup menu of each serial bus
analysis.
Undoing Auto Setup (Undo)
You can revert to the settings that were used immediately before you executed auto setup.