6-5
IM 701310-17E
Status Reports
1
2
3
4
5
6
App
Index
6.4 Extended Event Register
Reading the extended event register tells you whether changes in the condition register
(reflecting internal conditions) have occurred. A filter can be applied which allows you to
decide which events are reported to the extended event registe.
ACS
6
HST
7
5
4
3
2
1
0
PRN
TST
CAL
TRG
CUR
RUN
NGO
SCH
8
UME
9
MES
10
11
TEL
12
NSG
13
AN1
14
AN2
15
6
7
5
4
3
2
1
0
8
9
10
11
12
13
14
15
Condition register
:STATus:CONDition?
Transition filter
0
14
15
Extended event register
1
6
7
5
4
3
8
9
10
11
12
13
2
6
7
5
4
3
2
1
8
9
10
11
12
13
14
15
16
:STATus:FILTer<x>
{RISE
|
FALL
|
BOTH
|
NEVer}
:STATus:EESR?
FILTer<x>
→
The meaning of each bit of the condition register is as follows:
Bit 0
RUN (Running)
Set to 1 while waveform acquisition is in progress.
Bit 1
CUR (Cursor)
Set to 1 during cursor measurement.
Bit 2
TRG (Awaiting trigger)
Set to 1 when waiting for a trigger.
Bit 3
CAL (Calibration)
Set to 1 while calibration is in progress.
Bit 4
TST (Testing)
Set to 1 while self-test is in progress.
Bit 5
PRN (Printing)
Set to 1 while the built-in printer is operating, while data is
being output to an external printer (USB/network), or while
screen image data is being saved.
Bit 6
ACS (Accessing)
Set to 1 while a storage drive is being accessed.
Bit 7
MES (Measuring)
Set to 1 when automated measurement of waveform
parameters is in progress.
Bit 8
HST (History Search)
Set to 1 while history search is in progress.
Bit 9
UME (User Math Executing) Set to 1 while the user-defined computation is in progress.
Bit 10 NGO (Go/No-go)
Set to 1 while GO/NO-GO search is in progress.
Bit 11 SCH (Search)
Set to 1 while search is in progress.
Bit 12 TEL (Telecom Test)
Set to 1 while the telecom test is in progress.
Bit 13 NSG (N-Single)
Set to 1 while continuous acquisition is in progress when the
trigger mode is set to single (N).
Bit 14 AN1 (Analysis1)
Set to 1 while Analysis 1 is in progress.
Bit 15 AN2 (Analysis2)
Set to 1 while Analysis 2 is in progress.
The transition filter parameters detect changes in the specified bit (numerical suffix, 1 to
16) of the condition register in the following manner and overwrite the extended event
register.
RISE
The specified bit of the extended event register is set to 1 when the bit of the condition
register changes from 1 to 0.
FALL
The specified bit of the extended event register is set to 1 when the bit of the condition
register changes from 0 to 1.
BOTH
The bit of the extended event register is set to 1 when the bit of the condition register
changes from 0 to 1 or from 1 to 0.
NEVer
Always 0.