2-5
IM DL850E-01EN
Waveform Vertical Position (Vertical POSITION knob)
The DL850E/DL850EV can display the waveforms of the main analog-input channels, the waveforms of the
sub analog-input channels, and computed waveforms. Because the DL850E/DL850EV can display so many
waveforms, the waveforms may overlap and be difficult to view. If this happens, you can adjust the vertical
display position to make waveforms easier to view (vertical position).
The vertical position can be moved within the range of ±5 div.
When you change the vertical scale (V/div), the location of the vertical position mark does not change.
Use the POSITION knob to set the vertical position for each channel and sub channel.
The same POSITION knob is used to adjust the position of each channel and sub channel.
To change the position of a channel, press the key from CH1 to CH16 that corresponds to the channel.
To change the position of a sub channel, press an odd channel key from CH1 to CH15, and then press the soft
key that corresponds to the sub channel.
Position: 3 div
Position: –3 div
Position: 0 div
Vertical position
mark
• You cannot use the vertical POSITION knob to move the vertical position of a temperature, strain, or
computed waveform.
• When the menu for configuring all channels (ALL CH) or all sub channels (ALL Sub Channels Setup) is
displayed, you can use the job shuttle to change the vertical positions.
• When you change the vertical position, offset voltage, or upper or lower limit (when Variable is set to ON),
data that is outside of the measurement range is handled as overflow data. As shown in the figure below,
waveforms appear cut off when there is overflow data.
2 Vertical Axis