11.12 Changing the SCSI IDs
Function
SCSI IDs identify the devices connected to the SCSI bus. Each device on the bus must have a
unique ID. Default IDs are as follows.
SCSI ID (Own)
: 6
SCSI ID (External : external SCSI device) : 5
SCSI ID (Internal : internal hard disk)
: 4
* Internal hard disk and SCSI interface are options.
Range for SCSI ID numbers
0 to 7
Terminator ON/OFF
The terminator can be turned ON/OFF according to the system being used. The default setting is
“ON.”
If you are using this instrument by itself (not connected to anything else), turn the terminator
ON. Otherwise, the internal hard disk will not be recognized.
Precautions and Notes
• The same numbers cannot be set to SCSI ID (Own), SCSI (External) and SCSI (Internal).
• Do not set the SCSI ID number of the instrument (initial value 6) to the internal hard disk or
the external SCSI device.
• The SCSI device corresponding to the ID number of “External ID” is the device which
various data is saved to and read from. Set the ID number of the device you wish to save or
read the various data.
• When changing the SCSI ID numbers of the instrument and the internal hard disk, make sure
to press the “Initialize SCSI” soft key.
• If you change the ID for an external device at the instrument side, remember that you make
the same change at the external device’s side. For instruction on how to make the change at
the external side, refer to the documentation provided with you external device.
Operating Procedure
Changing SCSI ID for the instrument or for the Internal hard disk
1.
Press the
MISC
key.
2.
Press the “
SCSI
” soft key.
3.
Press the “
Own ID
” or “
Internal ID
” soft key to select the jog shuttle action.
4.
Turn the jog shuttle to set the ID.
5.
Press the “
Initialize SCSI
” soft key to execute the change of the ID number.
Select the SCSI device to access
3.
Press the “
External ID
” soft key so that the jog shuttle controls the “
External ID
.”
4.
Turn the jog shuttle to set the ID number of the SCSI device to save/load the various data.
Turn ON/OFF the terminator
Press the “
Terminator
” soft key and select “
ON
” or “
OFF
.”
→
Note
When changing the “External ID,” the “Initialize SCSI” soft key does not need to be pressed.