2.3 Displaying the Signals to Be Analyzed
Setting the HF Rejection
8.
Press the HF Reject soft key to select ON or OFF.
Setting the Hysteresis
9.
Press the Hysteresis soft key to select
or
.
Setting the Hold Off Time
10.
Press the Hold Off soft key.
11.
Turn the jog shuttle to set the hold off time.
You can move between the digits using the arrow keys. Pressing RESET resets
the value to 0.08
µ
s.
Explanation
Selecting the Trigger Source
To activate the trigger on the falling edge of the CS signal (CH4), set the trigger source
to CH4.
Setting the Trigger Level
Selectable range: 8 divisions within the screen
Resolution:
0.01 divisions (For example, the resolution for 2 mV/div is 0.02 mV.)
Setting the Trigger Slope
Select how the trigger source is to cross the specified level for activating the trigger from
the following three choices. To activate the trigger on the edge falling of the CS signal
(CH4), set the trigger slope to .
:
Trigger when the signal changes from below the trigger level to above the trigger
level (rising)
:
Trigger when the signal changes from above the trigger level to below the trigger
level (falling)
: Trigger on either the rising or falling edge.
Setting the Trigger Coupling
Select the trigger coupling from the following. To activate the trigger on the edge falling
of the CS signal (CH4), set the trigger coupling to DC.
AC: Uses a signal that is obtained by removing the DC component from the trigger
source signal.
DC: Uses the trigger source signal as-is.
Setting the HF Rejection
Specify “ON” if you wish to use a signal that is obtained by removing the high frequency
components (frequency components greater than 15 kHz) from the trigger source signal
as the trigger source.