1-31
IM 04L42B01-01E
Overview of Functions
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
• Recording Conditions of Event Data
Item
Description
Source channels
Same as the display data.
Sampling interval
Select from the available settings between 25 ms to 30 minutes.
However, you cannot specify an interval that is faster than the scan
interval.
File creation
A file is created when the specified data length is reached.
Files are also created in the following cases.
• When a file is created manually.
• When the memory sampling is stopped.
• When file creation is executed with the event action function.
•
After recovering from a power failure.
Mode
The available modes are
Free
(continuously record),
Single
, and
Repeat
. The recording operation varies depending on the mode as
follows:
Free
Press the START key to start recording (memory start) and the STOP
key to stop the recording (memory stop) .
Time
File
File
File
Adding data
Single
Pressing the START key places the DX in the trigger-wait state. When
the trigger condition is met, the DX records data for a specified time
(data length) and stops. From this point, the DX does not record even
if the trigger condition is met.
Time
File
Trigger condition met
Repeat
Pressing the START key places the DX in the trigger-wait state. When
the trigger condition is met, the DX records data for a specified time
(data length) and stops. The DX enters the trigger-wait sate again and
keeps recording the data for a specified time (data length) each time
the trigger condition is met. To stop the recording of the event data,
press the STOP key.
Time
File
File
Trigger condition met
Trigger condition met
Pretrigger of Event Data
The pretrigger can be specified in trigger mode.
This function is used to save the data before the point where the trigger condition
is met as event data. This function is convenient when you wish to record the data
before the occurrence of a certain event such as when an alarm occurs.
Specify the pretrigger as a percentage (0, 5, 25, 50, 75, 95, or 100%) of the recording
time (data length) of the event data. If set to 0%, the data after the trigger condition is
met is recorded.
Time
Trigger condition met
Data before
the trigger
Data after
the trigger
Trigger for Starting the Event Data Recording
When set to trigger mode, you can set various conditions for starting the recording.
Example: Key operation, alarm occurrence, specific time, or remote control
1.4 Data Storage Function