5-23
IM CW500-01EN
Flicker Filter Coefficient Settings
Filter Coefficient
Select an appropriate filter coefficient according to the nominal voltage.
To accurately measure flicker, the nominal voltage, nominal frequency,
and filter coefficient must be set properly for the actual DUT. If possible,
use the same voltage for the nominal voltage and filter coefficient.
Settings
230V*/220V/120V/100V
* Default value
<Procedure>
Use the up and down arrow keys to move to Filter.
→
Press ENTER to
show a pull-down menu.
→
Select the appropriate filter coefficient.
→
Press ENTER to confirm or ESC to cancel.
5.3 Measurement Setup