2-2
IM CW500-01EN
2.2 Connectors
AC voltage input terminals
(VN, V1, V2, V3)
Current input terminals
(A1, A2, A3, A4)
Terminal covers
Power inlet
Wiring System
AC Voltage
Input Terminal
Current Input
Terminal*
Single-phase two-wire (1 system) 1P2W×1 VN, V1
A1
Single-phase two-wire (2 system) 1P2W×2 VN, V1
A1, A2
Single-phase two-wire (3 system) 1P2W×3 VN, V1
A1, A2, A3
Single-phase two-wire (4 system) 1P2W×4 VN, V1
A1, A2, A3, A4
Single-phase three-wire (1 system) 1P3W×1 VN, V1, V2
A1, A2
Single-phase three-wire (2 system) 1P3W×2 VN, V1, V2
A1, A2, A3, A4
Three-phase three-wire (1 system) 3P3W×1 VN, V1, V2
A1, A2
Three-phase three-wire (2 system) 3P3W×2 VN, V1, V2
A1, A2, A3, A4
Three-phase three-wire 3A
3P3W3A V1, V2, V3
A1, A2, A3
Three-phase four-wire
3P4W×1 VN, V1, V2, V3 A1, A2, A3
* Current terminals that are not used in wiring can measure only rms values
and harmonics.