6-24
IM CW500-01EN
Changing the Displayed Harmonic Orders
Use the up and down arrow keys to vertically scroll the screen thereby
changing the displayed orders.
Graph/List
Press F1 (graph/list) to display voltage, current, power harmonics,
from 1st to 50th order, in bar graphs. On the bar graph screen, only the
percentage content is displayed.
Percentage Content, Phase Angle, RMS Value (Power)
Press F2 (percentage content/phase angle/rms value (power) to change
the measurement items displayed in the list. If V (voltage) or A (current) is
selected, select percentage content, phase angle (in reference to V1 (0°)),
or rms value. If P (ΣP) (power) is selected, select percentage content,
phase angle (current-to-voltage difference between channels), or power.
V/A/P/ΣP
Press F4 (V/A/P/ΣP) to select the harmonic analysis parameter (voltage,
current, power, or system total or total sum power).
6.6 Harmonics