<8. Specifications>
8-5
IM 80J01A01-01E
8.3
Measurement sample vessels
(1) Microplate
Following microplate types are applicable
・
6 well
・
12 well
・
24 well
・
48 well
・
96 well
・
384 well
※
1
・
1536 well
※
1
(2) Dish
Following dish types are applicable
・
35 mm Dish (By using 35 mm dish sample holder, up to
3 dishes are measurable in one experiment)
・
60 mm Dish (By using 60 mm dish sample holder, one dish
is measureable in one experiment)
(3) Slide glass
By using slideglass sample holder, up to 4 slide glasses are
applicable in one experiment
(4) Cover Glass Chamber
By using cover glass chamber holder, one cover glass chambdr is
applicable in one experiment
※
1
If 384 well microplate is measured by 2x objective lens, multiple wells are measured
by one FOV. Therefore in this case, measurement is inhibited. (Warning is
displayed.)
If 1536 well microplate is measured by 2x/ 4x objective lens, multiple wells are
measured by one FOV. Therefore in this case, measurement is inhibited. (Warning
is displayed.)
For details of sample vessels usable in CQ1, please refer to Technical
Information (TI 80J01A01-01E).
Summary of Contents for CQ1
Page 2: ...This document corresponds to the following versions of software Measurement software R1 04 ...
Page 8: ...vi IM 80J01A01 01E Right Side of Utility Box Aperture label ...
Page 24: ......
Page 34: ......
Page 48: ......
Page 177: ... 5 Functions of CQ1 Software 5 81 IM 80J01A01 01E Drag tab to move the window ...
Page 236: ......
Page 238: ......
Page 252: ...7 14 7 Trouble Shooting IM 80J01A01 01E Set sub cover and tighten 2 screws ...
Page 256: ......
Page 266: ......