4-2
4.2 Error Codes, Error Messages, and Actions
Number
Error description
Cause of Error
Corrective Action
0
No error
—
—
11
Command error
An undefined command was received
(USB communication).
Check the command.*
12
Parameter error
Command parameter designation is
incorrect (USB communication).
Check the command.*
13
Execution error
A command that cannot be executed
in the current condition was received
(USB communication).
Check the command.*
16
Adjustment error
An error was detected during
adjustment.
Servicing is required.
17
Zero calibration error Zero calibration error.
Check the installation orientation of
the CA700.
20
Source error
24 V LOOP measurement power
supply error
Check the connection.
23
Source error
Excessive current or voltage was
detected in the source output.
Check the connection.
24
Source error
Insufficient current supply
Replace the batteries.
33
Storage error
The data storage destination is full.
Data cannot be saved.
Check the save destination.
34
Storage error
Data cannot be saved due to some
error occurring at the data storage
destination.
Check the save destination.
50
Measurement error
Pressure sensor error
Servicing is required.
51
Measurement error
Tilt sensor error
Servicing is required.
52
Measurement error
A/D converter error
Servicing is required.
* See the communication commands (section 5.3) in IM CA700-01EN, which is included in the
accompanying CD.
Error message
Symptom
Action
A communication error of the
external pressure sensor.
Reconnect the sensor.
A communication to the PM100
cannot be established
•
Check the external pressure sensor icon
•
Reconnect the PM100
•
Power on the main unit again
•
Replace the connection cable
If the error cannot be resolved, a service is
required.
A memory error of the external
pressure sensor.
Reconnect the sensor.
The memory inside the PM100 is
faulty
•
Reconnect the PM100
•
Power on the main unit again
If the error cannot be resolved, a service is
required.
A measurement error of the
external pressure sensor.
Reconnect the sensor.
Pressure cannot be measured
correctly
•
Reconnect the PM100
•
Power on the main unit again
If the error cannot be resolved, a service is
required.
The external sensor has been
changed.
Reboot the device.
Changed to an external pressure
sensor with a different range while
the instrument was on.
Press ESC or ENTER, and then restart the
instrument.
The measured pressure value is
“------”
·
The connection cable of the
PM100 is not connected properly
·
The PM100 is faulty
•
Check the external pressure sensor icon
•
Reconnect the PM100
•
Power on the main unit again
If the error cannot be resolved, a service is
required.