3-12
IM CA700-02EN
3.3 Calibrating the Pressure Transmitter
Procedure
Selecting the Calibration Procedure
1.
In the calibration display, press
Select Proc. (F2)
.
The calibration procedure setting display appears.
An asterisk is displayed next to the selected calibration procedure.
2.
Use the ▲ or ▼ key to select a calibration procedure,and press
ENTER
. Here, select the
calibration procedure whose Source is pressure (P) and Measure is mA Loop, as set in section
3.2.
The calibration procedure is selected, and the calibration display returns.
ENTER
The information of the selected
calibration procedure appears.
Saving Calibration Data at Calibration Points
3.
Press
As Found (F3)
.
The display switches to the source and measure value display.
Pressure transmitter output
(loop current)
Pressure transmitter input
(pressure)
Calibration point (current calibration
point/total number of calibration points)
Pressure transmitter input value at the calibration point
Error in the pressure transmitter output