3-8
IM CA700-01EN
Setting Sensor Type and Range (for external pressure sensor)
The Sensor Type is fixed to 16 MPa.
The procedure for setting Range is the same as in pressure measurement. See section 2.1.
Setting Unit, Averaging, and Scaling (Pressure)
These items are the same as with pressure measurement. See section 2.1.
Setting the 0% Value and 100% Value
17.
Set values that correspond to 0% and 100% of the measured values you set in steps 5 and 6.
Use the ▲ or ▼ key to select
0% Value
, and press
ENTER
.
18.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
Likewise, set the 100% value.
Setting the Number of Calibration Points (Number Of Points)
19.
Use the ▲ or ▼ key to select
Number Of Points
, and press
ENTER
.
20.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
The range is 1 to 10.
Setting the Calibration Direction (Direction)
21.
Press the ▲ or ▼ key to select
Direction
.
The options appear on the function menu.
0% to 100%
100% to 0%
0% to 100% to 0%
22.
Press
Up (F1)
,
Down (F2)
,
or Up/Down (F3)
.
Confirming the Settings
23.
After entering the information, press
Setting Done (F5)
. The settings are confirmed.
Press
ESC
to return the display.
3.3 Setting the Device Information and the Calibration Conditions of the Pressure Transmitter