2.4 Measuring DC Current and DC Voltage
Procedure
Refer to the Getting Started Guide (IM CA700-02EN), and correctly connect the CA700 to the device
under measurement.
DC current and DC voltage measurement is valid when the pressure measurement of Display1 is set
to normal pressure measurement (P). Follow the procedure in section 2.1 to set Function to normal
pressure (P).
1.
From the top menu, use the ▲ or ▼ key to select
Measure and Source
, and press
ENTER
.
The source/measure screen appears.
Display1
Display2
Switches to the MEASURE display
(MEASURE mA, MEASURE mA LOOP, MEASURE V )
2.
When Display2 is showning the source screen, press
SOURCE/MEASURE
to switch Display2
to MEASURE.
3.
Press
Setting (F5)
.
The setting categories are displayed.
Setting category menu
4.
Press
Display2 Setting (F2)
.
The measurement function setting screen appears.
Turns averaging on and off
Turn scaling on and off and set scaling.
Turn the alarm on and off and
set the alarm output status display.
Turns the percentage display on and off
Sets the relative display based on the reference value
Initializes the settings
Confirms the settings
Set the measurement function.
Setting Measurement Functions
5.
Use the ▲ or ▼ key to select Function, and press ENTER.
The measurement function options are displayed.
6.
Use the ▲ or ▼ key to select the measurement function, and press
ENTER
.
Setting Measurement Conditions
If necessary, set the measurement conditions.
The procedure for setting averaging, scaling, alarm, percentage display, and relative display items and
for initializing the settings are the same as in pressure measurement. See section 2.1.
2-18
IM CA700-01EN