Setting the Source Value Sweep Method
18.
Press the ▲ or ▼ key to select
Step Type
.
Sweep methods appear on the function menu.
19.
Press
Linear (F1)
or
Step (F2)
.
Confirming the Settings
20.
After entering the measurement conditions, press
Setting Done (F5)
.
The screen shown in step 1 appears.
Sweeping Automatically
21.
After step 19, press
Step/Sweep (F1)
.
22.
Press
Sweep (F1)
.
23.
Press
Source On (F1)
.
The F1 key changes to “Source Off,” the output status display changes to SOURCE ON, and the CA700
sweeps the source values according to the sweep method settings.
Stops sweeping
Starts data saving
Pauses sweeping
Pausing and Resuming Sweeping
24.
The F2 key displays Sweep Pause. To pause sweeping, press
Sweep Pause (F2)
.
The F2 key changes to Sweep Start, and the output value is held. The CA700 continues to generate current
or voltage.
To resume sweeping, press
Sweep Start (F2)
. The CA700 will continue sweeping from where it
paused.
Saving Data
25.
Press
Save (F3)
or
Logging Start (F3)
. The source value is saved along with the pressure
measurement data.
What appears at F3 varies depending on the data save method.
Note
For details on setting the data save method (Save or Logging), see section 4.1.