2-8
IM CA700-01EN
2.2 Measuring
Leak Rates
Procedure
WARNING
To prevent injury to the user and damage to devices caused by incorrect connection or
pressure leakage, be sure to connect tubing correctly to the CA700 pressure input port by
following the instructions in the Getting Started Guide, IM CA700-02EN.
Selecting the Measurement Function
1.
From the top menu, use the ▲ or ▼ key to select
Measure and Source
, and press
ENTER
.
The source/measure screen appears.
Display1
Display2
2.
Press
Setting (F5)
.
Display1 Setting and Display2 Setting appear on the function menu.
Function menu
3.
Press
Display1 Setting (F1)
.
The pressure measurement setting screen appears.
4.
Use the ▲ or ▼ key to select
Function
, and press
ENTER
.
The measurement function options are displayed.