7-4
IM CA700-01EN
7.2 Adjusting the Measurement Function
Procedure
To adjust the CA700, connect a DC voltage/current generator to the CA700 voltage or current
measurement terminals. Use a DC voltage/current generator with sufficient specifications to adjust the
CA700.
CA700
L
H
CA700
L
H
DC voltage
DC current
Pressure switch resistance
CA700
DC voltage/current source
DC voltage/current source
L
H
Reference resistance
mA loop
CA700
L
H
DC voltage/current source
Absorption of 24 V, 24 mA
Digital
multimeter
L
H(A)
1.
While holding down
HOLD
and the backlight (
) key, turn on the CA700. Keep holding the
keys down until the self-test screen appears.
The User Adjustment screen appears.
Adjust the DC voltage measurement.
Adjust the DC current measurement.
Adjust the loop power supply current measurement.
Adjust the pressure switch resistance measurement.
Adjust pressure measurement.
(Contact your nearest YOKOGAWA
dealer or service.)