5-4
IM CA500-01EN
5.2 Saving Sweeps
Procedure
1.
Set DATA SAVE to ON in the step sweep or program sweep settings according to section 2.9,
2.
Execute a sweep. After the sweep is completed, the source values and measurement values of
each step are saved.
Description
When a step sweep or program sweep is completed, the source values and measurement values
of each step are saved automatically.
CA500
The saved information is the same as that explained in “Saving Data Manually” and Section 5.1.
A total of up to 100 data entries can be saved including manually save data.
Depending on the number-of-steps setting, 100 data entries may be exceeded when a sweep is
saved. In this situation, an error message is displayed when a sweep is started.
Data is saved in a dedicated format of this instrument.
If you abort a sweep, the data up to that point is saved.