5-1
IM CA500-01EN
1
2
3
4
5
6
7
Index
Saving Data
5.1 Saving Data Manually
Procedure
1.
While sourcing or measurement is in progress, press
SAVE
. The source value and
measurement value are saved to the internal memory of this instrument when SAVED is
pressed.
Description
The method to save data is different between the CA500 and CA550.
CA500
You can save the date and time, the selected function, range, measurement value, and source
value when SAVE is pressed.
Number of Data Values That Can Be Saved
A total of up to 100 data entries can be saved including the saved sweep data.
Memory numbers from 001 to 100 are automatically assigned to the data entries.
Saving is not possible beyond 100 data entries. Delete unneeded data.
Saves up to 100 data entries
List of saved data
Data Format
The format that data is saved in is exclusive to this instrument.
You can load and display the data on this instrument. For instructions on how to load data, see
section 5.3, “Loading and Deleting Saved Data”.
Data can also be loaded into a PC using communication commands.
Saved Information
The following information is saved.
Function1 Information
Saved data
Notes
Measured value
Function
Range
0% value
100% value
Contact input setting
Count time
Chapter 5 Saving Data